Precision and Accuracy of Oxygen Flowmeters: A Systematic Review with Meta-Analysis

Authors

  • Caio Henrique Veloso da Costa Salvus

DOI:

https://doi.org/10.36557/pbpc.v4i2.373

Keywords:

Oxygen Therapy, Flowmeters, Precision, Accuracy, Patient Safety, Health Technology

Abstract

Introduction: Oxygen therapy is a ubiquitous pharmacological therapy, whose efficacy and safety critically depend on the accuracy of the delivery device. However, the accuracy of conventional oxygen flowmeters is often questioned. Objective: To conduct an in-depth systematic review with meta-analysis to quantify the precision and accuracy of oxygen flowmeters, identify the drivers of inaccuracy, and outline the implications for clinical practice and health policy. Methods: A systematic search of the PubMed, EMBASE, Cochrane Library, and LILACS databases was conducted for studies published between 1980 and 2025 that evaluated the accuracy of oxygen flowmeters against a gold standard. The meta-analysis was conducted using a random-effects model to calculate the weighted mean difference (WMD) of the percentage deviation. Subgroup analyses were performed by technology (analog vs. digital), flow range, and presence of pressure compensation. Results: Fifteen studies were included, totaling 1,847 flowmeters and over 15,000 measurements. The global meta-analysis revealed a weighted mean deviation of 8.7% (95% CI: 5.2-12.1%), with extreme heterogeneity (I² = 89%). Individual deviations varied dramatically, from -52% to +85% of the nominal value. Analog flowmeters exhibited a linear deterioration in accuracy, with median deviations increasing from 2% at 1 L/min to over 30% at 15 L/min. In contrast, digital flowmeters maintained a consistently low deviation (<1.5%) across the entire operating range (p < 0.001). Inaccuracy was significantly higher at low flows (<3 L/min) and in devices without pressure compensation. Conclusion: The evidence conclusively demonstrates that analog oxygen flowmeters lack the precision necessary for safe therapy.

Downloads

Download data is not yet available.

References

O'Driscoll BR, Howard LS. Audit update: British Thoracic Society emergency oxygen audits. Thorax. 2011;66(8):734-5.

Arora N, Dennis A, Willson J, Norrie J, Tunstall M. Delivery of oxygen by standard oxygen flowmeters. Anaesthesia. 2021;76(11):1546-7.

Austin MA, Wills KE, Blizzard L, Walters EH, Wood-Baker R. Effect of high flow oxygen on mortality in chronic obstructive pulmonary disease patients in prehospital setting: randomised controlled trial. BMJ. 2010;341:c5462.

Cairo JM. Administering medical gases: regulators, flowmeters, and controlling devices. In: Cairo JM, Pilbeam SP, editors. Respiratory care equipment. 6th ed. New York: Mosby's Inc.; 1999. p. 66.

Davidson J, Gazzeta C, Torres LC, Jardim JR, Nascimento OA. Precision and accuracy of oxygen flow meters used at hospital settings. Respir Care. 2012;57(7):1071-5.

Howard WR. Flow meters: what do you mean the flow rate from one does not equal the same flow rate from another? Respir Care. 2012;57(7):1197-8.

Ehrenkranz RA, Walsh MC, Vohr BR, Jobe AH, Wright LL, Fanaroff AA, et al. Validation of the National Institutes of Health consensus definition of bronchopulmonary dysplasia. Pediatrics. 2005;116(6):1353-60.

Shimoda LA. Hypoxic regulation of ion channels and transporters in pulmonary vascular smooth muscle. Adv Exp Med Biol. 2010;661:221-35.

O'Driscoll BR, Howard LS, Davidson AG. BTS guideline for emergency oxygen use in adult patients. Thorax. 2008;63(Suppl 6):vi1-68.

Costa CHV, Tavares FCCJ, Silva Júnior MG. Precisão de fluxômetros de oxigênio analógicos e digitais: estudo comparativo em condições padronizadas de pressão. Rev FT. 2025;29(147):30-42.

Duprez F, Barile M, Bonus T, Cuvelier G, Ollieuz S, Mashayekhi S, et al. Accuracy of medical oxygen flowmeters: a multicentric field study. Health. 2014;6(15):1978-83.

Derom E, Meijer EJ, Van Enschot JWT. An on-demand oxygen flow meter for enhanced patient comfort and reduced oxygen cost in hospitals. COPD. 2022;19(1):274-81.

Pagliocchi A, Davidson J, Gazzotti M, Oliveira A, Nascimento O, Jardim J, et al. Accuracy and precision of oxygen flowmeters used in neonatal and pediatric care. Eur Respir J. 2019;54(Suppl 63):PA1231.

Murphy M, Hodgson D, Bello N. On accuracy and precision of flowmeters used for oxygen therapy in a veterinary teaching hospital. Vet Anaesth Analg. 2018;45(1):41-47.

Fissekis S, Hodgson D, Bello N. Effect of cleaning status on accuracy and precision of oxygen flowmeters of various ages. Vet Anaesth Analg. 2017;44(4):890-8.

Published

2025-07-22

How to Cite

DA COSTA, Caio Henrique Veloso. Precision and Accuracy of Oxygen Flowmeters: A Systematic Review with Meta-Analysis. Periódicos Brasil. Pesquisa Científica, Macapá, Brasil, v. 4, n. 2, p. 114–121, 2025. DOI: 10.36557/pbpc.v4i2.373. Disponível em: https://periodicosbrasil.emnuvens.com.br/revista/article/view/373. Acesso em: 27 sep. 2025.